Gold APP Instruments
Corporation China had coordinated with ISU (Iowa State University) to
established one gas adsorption demo lab on campus of ISU. Professor Mufit Akinc
from the Materials Science & Engineering said they had already well
installed and calibrated all demo analyzers (gas pycnometer density analyzer
G-DenPyc 2900 & BET surface area and porosimetry analyzer V-Sorb 2800P)
with helps from GAPP engineer, were ready to receive visitors to go this lab to
view analyzers onsite, to test their samples and value instruments quality.
BET Surface Area|Porosity Analyzer|Gas Pycnometer Helium True Density Equipment|High Pressure Volumetric Analyzer| Langmuir/BET/Gibbs/BJH/MP/SF/DR/DA/HK/T-Plot Analyzer
volumetric high pressure gas sorption analyzer H-Sorb 2600 for methoane analysis
2016年3月22日星期二
2016年1月7日星期四
Specific Surface Area and Porosity Analyzer Exported to Canada Manson Environmental Corporation
Gold APP Instruments
Corporation China had export one set volumetric specific surface area and
porosity analyzer to Canada catalyst manufacturer--Manson Environmental
Corporation. This is the first set Chinese brand analyzer came into Canada
market.
Manson Environmental
Corporation is an inorganic chemicals manufacturer in Oakville. This private
company was founded in 1986.
Gold
APP Instruments China also produce gas pycnometer analyzer for true density,
the open and closed pore cells testing; high pressure volumetric adsorption
analyzer for PCI, TPD, adsorption/ desorption isotherms testing. Welcome email
to sales@jinaipu.com for more details.
2014年4月2日星期三
2012年12月19日星期三
automated determination of high pressure and high temperature gas sorption analyzer H-Sorb 2600
H-Sorb 2600 high pressure gas sorption parameters
Analysis Method: static
volumetric, high pressure gas sorption
Versatility: adsorption/desorption
isotherms measurement (normal to 500℃), Gibbs supercritical adsorption measurement
Data Reduction: Langmuir
model regression isotherm, Langmuir maximum adsorption constant L and
adsorption pressure constant B measurement; Langmuir correction model loading-ratio
correlation regression (LCR) isotherm; Ono-Kondo (OK) lattice model isotherm
regression; adsorptive gas density and vapor-liquid equilibrium accuracy
calculation under different pressures and temperatures
Measuring Ranges※: continuous adsorption and desorption measurement from
normal to 200Bar pressure
Accuracy: repeatability
errors <3%
Temperature Range: normal
to 500℃,
accuracy±0.1℃
Sample Ports: two
samples analyzing and two degassing, stainless-steel micro welding sample
cells, software integrated temperature PID adjustment
Transducer Accuracy: imported
high precision transmitter, accuracy can reach 0.05% of F.S. (full-scale),
long-duration usage stability is 0.025% of F.S.
Ultimate Vacuum: 4x10-2
Pa (3x10-4 Torr), molecular pump system (1x10-6 Pa) is
optional
Adsorbate Gas: high
purity nitrogen, CO2 (≥99.999%) or
others (Ar, Kr, H2, CH4 etc.)
Data Acquisition:
high-precision and high integration data acquisition modules, minimal error,
strong anti-interference ability
Specifications: Height
22.05 inches (56cm)* Width 20.87 inches (53cm)*Depth 23.62 inches (60cm); 132
pound (60kg); 220V/5A
H-Sorb 2600 high pressure gas sorption features:
Control System※: imported VCR interface high pressure pneumatic valve can realize auto on-off within 200Bar, sealing performance is top to 1x10-10 Pa.m3/s, can be used more than 500 million times; programmable logic controller (PLC) system, high integration and sound anti-interference, improve stability and prolong life
Analysis Management※: parameters be set by software, fully automated, unattended operation at night; H-Sorb mode is available for accurate control gas charging pressure points, gain ideal data points
Manifold System: imported 316L stainless steel thick-walled manifolds, micro welding sealing can reduce dead space largely; metal VCR connection is convenient for installation and uninstallation
Safety Measures※: H-Sorb asymptotic technology realizes auto gas-charging and degassing, prevents high pressure dangers caused by artificial misoperation, and also minimizes large differential pressure’s impact to pressure transducer
H-Sorb 2600 high pressure gas sorption application
1.
Study and research in high temperature and
pressure gas sorption, supercritical gas performance, microporous materials,
hydrogen storage materials, coalbed gas, oil exploration, methane etc.
2.
Catalysts, molecular sieves, active carbon,
carbon nano-tube, fuel cell, hydrocarbon and other materials
Currently, dealers, distributors, agents, wholesalers are recruiting. If
you interest or deem your country has a big and unexplored market, pls. let us
know. Thanks!
For more details, pls. contact us:
Web: www.jinaipu.com
Email: sales@app-one.com.cn
Tel.: 86-10-82133318 #810
Mobile: 86-18210009838
Skype: Gold-APP-Instruments
订阅:
评论 (Atom)